7-22 QX Hawk Industrial Imager User Manual
Calibration
Calibration
Sending the <@CAL> command from ESP’s Terminal initiates auto-calibration without
entering the auto-calibration menu. The calibrated parameters are determined by the
settings of the Calibration Options command.
Lens Calibration
Important:
The
<@OPTIC>
command should only be used when changing the lens module.
Important: <@OPTIC> is not supported by the QX Hawk C-Mount Imager.
Sending the <@OPTIC> command puts the imager into a setup mode for the purpose of
improving the accuracy of the focal distance parameter (<K525>). The imager comes with
a default focal curve based on unit averages. When this isn’t close enough to obtain all
distances, the <@OPTIC> command can be used to start a 2-point test to normalize the
focal distance and improve accuracy. This should generally be done with lenses that were
not originally shipped with the imager.
Important: <K525> is not supported by the QX Hawk C-Mount Imager.
The table below shows the setting and test requirements to improve distance parameter
accuracy.
Lens Type and Command Inside Test Requirements Outside Test Requirements
<K525,,1> (.) UHD 15°
10 mil/0.25 mm Data Matrix @
1.0”/25 mm
30 mil/0.75 mm Data Matrix @
10”/254 mm
<K525,,2> (..) HD 30°
10 mil/0.25 mm Data Matrix @
1.0”/25 mm
30 mil/0.75 mm Data Matrix @
10”/254 mm
<K525,,3> (…) SD 45°
10 mil/0.25 mm Data Matrix @
1.0”/25 mm
30 mil/0.75 mm Data Matrix @
10”/254 mm
<K525,,4> (
) SD 12°
10 mil/0.25 mm Data Matrix @
1.0”/25 mm
30 mil/0.75 mm Data Matrix @
10”/254 mm