8-36
EZ Match User Manual
Configurable Output 1
Number of Triggers
Number to Output On
Decodes/Trigger Threshold
Definition: The number of triggers to examine for the trend analysis condition.
Serial Cmd:
<
K780,
trend analysis mode,
number of triggers
,number to output on,
decodes/trigger threshold>
Default: 0
Options: 0 to 255
Usage: Example: If Number to Output On is set to 3 and Trend Analysis Mode is
set to No Read, then the output will not be activated until 3 No Reads have
occurred.
Definition: Sets the number of Trend Analysis Mode events (mismatches, No Reads,
or reads/trigger as configured by Trend Analysis Mode) to occur within the
trigger evaluation period before activating the associated output.
Serial Cmd:
<
K780,
trend analysis mode,number of triggers,
number to output on
,
decodes/trigger threshold>
Default: 0
Options: 0 to 255
Definition: When set to this mode, and when the appropriate output is set to output on
trend analysis, the imager will function in a
Decodes per Trigger
mode during
the read cycle and the trend analysis operation. Output will be activated
based
on whether or not the symbol decode count at the end of the read cycle
crosses the decodes per trigger threshold.
Note: Although this setup causes the imager to function in a Decodes per
Trigger mode, the decode count will only be appended to the symbol data if
the status of the Decodes per Trigger command is enabled.
Serial Cmd:
<
K780,
trend analysis mode,number of triggers,number to output on,
decodes/trigger threshold
>
Default: 0
Options: 0 to 65535