Chapter 7 OMR
MS-9 Scanner User’s Manual 7-13
Auto Frame Status
Average Bar Width
Average Cell Width
Leading Margin
Definition: When Enabled, OMR calibration command will adjust laser framing to
frame OMR calibration bar code with sufficient quite zones. When Auto
Frame is enabled, the OMR calibration symbol must be in the center of
the scan for a successful decode to occur.
When Disabled, OMR calibration command will not adjust laser framing.
Serial Cmd: <KPcalibrate delay status,auto frame status,avg bar width,avg cell
width, leading margin,scan count value>
Default: Enabled
Options: 0 = Disabled 1 = Enabled
Definition: Determined by OMR calibration. It represents the average value of the
bar widths from calibrated OMR symbol.
Usage: READ ONLY.
Serial Cmd: <KPcalibrate delay status,auto frame status,avg bar width,avg cell
width, margin,scan count value>
Definition: Determined by OMR calibration. It represents the average of the field
widths from calibrated OMR symbol. Count resolution is 15.25 nS.
Usage: READ ONLY.
Serial Cmd: <KPcalibrate delay status,auto frame status,avg bar width,avg cell
width,leading margin,scan count value>
Definition: Determined by OMR calibration. It represents the count value from the
start of the scan to the gate bar of OMR symbol. Count resolution is
15.25 nS.
Usage: READ ONLY.
Serial Cmd: <KPcalibrate delay status,auto frame status,avg bar width,avg cell
width,leading margin,scan count value>